![PHILIPS Respironics DreamStation BiPAP AVAPS Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/philips-respironics/dreamstation-bipap-avaps/dreamstation-bipap-avaps_service-and-technical-reference-manual_1550628072.webp)
P
AGE
72
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Code
Error Name
Error Level
Failed
Component
Actions
E-225
ERR_CRITICAL_HIGH_P
RESSURE
STOP
1. Air Path
2. PCA
3. Blower Box
1. If applying negative
pressure to the unit remove
that first and retest.
2. Load latest software
version if not up to date.
3. Clear air path including
filter and test.
4. If test fails, replace PCA
and retest.
E-226
ERR_SENSOR_PRESS_
UNRELIABLE
STOP
1. Air Path
2. PCA
3. Blower Box
1. If applying negative
pressure to the unit remove
that first and retest.
2. Load latest software
version if not up to date.
3. Clear air path including
filter and test.
4. If test fails, replace PCA
and retest.
E-227
ERR_LOW_MOTOR_RP
M
CONTINUE
1. Air Path
2. PCA
3. Blower Box
1. If applying negative
pressure to the unit remove
that first and retest.
2. Load latest software
version if not up to date.
3. Clear air path including
filter and test.
4. If test fails, replace PCA
and retest.
E-230
ERR_HUMID_PLATE_CY
CLE_OVERRUN
REBOOT
PCA
1. Clear error log and test.
2. Test humidifier
separately.
E-231
ERR_HUMID_PLATE_NO
_MEASUREMENT
REBOOT
1. Heater Plate
2. PCA
1. Clear error log and test.
2. Test humidifier
separately.
E-240
ERR_PACKET_CREATIO
N
CONTINUE
N/A
1. Clear error log and test.