P
AGE
101
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Replacing the Flow and Pressure Sensor Seals
Kit
Included in Kit
Tools Required
Flower Sensor Seal
•
Flower Sensor Seal
•
Torx Screwdriver (T10 is
recommended)
Pressure Sensor Seal
•
Pressure Sensor Sea
To install the Seals:
1. Remove all components as instructed in the previous sections.
2. Lift the Seals away from the Blower Box Assembly.
Note:
The Pressure Sensor Seal may remain attached to the Pressure Sensor on the PCA. Remove it from the
PCA if this is the case.
To install the Seals:
1. Align the Seals into the Blower Box Assembly grooves. Ensure the seals are flush against the Blower
Box Assembly.
2. Assemble the remainder of the device as instructed in the previous sections.
Pressure Sensor Seal
Flow Sensor Seal