4. MANUAL MODE
ZODIAC 901 features three dedicated pushbuttons,
Tymp.
,
Reflex
and
Threshold
, enabling the user to access any of the corresponding three
test modes from any other function or part of the instrument.
Setup
From any test in Manual Mode, you may press
Setup
to access the
Measurement Setup sub-menu, thus enabling you to change whichever
test parameters are valid at that given time. The relevant parameters are
listed after the description of each test given below.
Setup Inactive
During Test!
NOTE:
The
Setup
pushbutton is inactive while a test is in progress, i.e.
nothing will happen if this button is pressed during a test.
Pressing one of the three Manual Mode pushbuttons accesses a sub-
menu when Advanced Mode has been selected from the Main Menu,
enabling the user to select the desired test. This Chapter describes how
to perform each of the range of tests available from these menus.
Alt. Select
If it is required to erase the current test data, smooth a tymp. curve (Car-
tesian graph, not available with Version 2.
x
of the ZODIAC 901), access
the marker function, transmit data to an external computer, register a
change of ear, or adjust the contrast of the Liquid Crystal Display, press
Alt. Select
at any time except while a measurement is in progress (see
§5.1).
Air Release
The ZODIAC 901 Middle-Ear Analyzer incorporates a pump which is
used to apply a small static pressure or depression to the patient’s ear
canal. Although this pressure is no more than would be experienced by
ducking one’s head just below the surface of the water in a bath, it may
occasionally be sufficient to cause unexpected distress to a patient with
an ear disorder. The pressure may be relieved promptly by pressing the
Air Release
button, which automatically resets the pump position.
Most of the parameters used in the following tests which can be set by
pressing a
Softkey
repeatedly can also conveniently be adjusted by
holding the same
Softkey
down and turning the
Rotary Knob
in either
direction.
Operation Manual
29
GN Otometrics
MANUAL MODE
Chap. 4
ZODIAC 901 Middle-Ear Analyzer
Содержание Madsen ZODIAC 901
Страница 1: ...901 Middle Ear Analyzer User Manual Doc no 7 25 1800 11 Part no 7 25 180 0459 ...
Страница 6: ......
Страница 100: ...94 Operation Manual ...