Setup Parameters
Setup parameters available in the Air Conduction Threshold Mode (see
also §5.2.3.3) are:
Intensity Level increment: 1, 2, 5 or 10 dB
Pump: on or off
Pump On
If the pump is on, pump pressure can be increased or decreased manu-
ally by holding down
Softkey 5
and turning the rotary knob to the left
or right. Pressure in daPa can be read out from the display.
When the pump is enabled, the pressure range is changed to extended
range (-600 daPa to +400 daPa).
Gellé’s test
The Air Conduction Threshold mode may also be used to perform
Gellé’s test, using the probe to apply a static pressure or depression to
the ear canal (in Air Conduction Threshold mode there is no probe tone).
Remember to disconnect the TDH 39 headphone from the miniature jack
in the side of the probe, if connected.
An audiometer is required to perform audiometry with bone conduc-
tion.
Apply the bone conductor to the patient at the mastoid. Record the pati-
ent’s audiogram several times with different pump pressures. Lack of
variation of threshold between the different audiograms indicates a
probable fixation of the
stapes
.
52
Operation Manual
MANUAL MODE
Chap. 4
GN Otometrics
THRESHOLD TESTING
4.3
ZODIAC 901 Middle-Ear Analyzer