An example of a simple program sequence is given below (in Advanced
Mode). This sequence performs an automatic tymp. sweep on each ear
in turn, with A.F.A.P. pump speed and auto scaling, prints the two tym-
panograms out, and advances the patient counter. One pause is in-
cluded, to allow the operator to insert the probe into the patient’s other
ear. Execute the following keystrokes from the Program Mode Select
menu, see Fig.7.2:
Softkey 1 (Test1)
Tymp.
Softkey 2 (Sweep)
Setup
Softkey 1 (Pump Speed: A.F.A.P)
Softkey 3 (Scaling: 4.5 ml)
Softkey 3 (Scaling: Auto)
Setup
Softkey 2 (Start)
Standby
Alt. Select
Softkey 4 (L/R)
Softkey 2 (Start)
Printer
Softkey 3 (Print All)
Alt. Select
Softkey 4 (L/R)
Main Menu
Softkey 1 (New Pt.)
Softkey 1 (Yes)
Auto Test
Softkey 5 (End Program)
Enter Test Name (use rotary switch)
Auto Test
Softkey 1 (save user test)
Power instrument off (to disable programming mode)
Auto Test
Now power instrument on again, press
Auto Test
and
Softkey 1
to start
your own User Test.
Use Default Config.
NOTE: Since existing test parameters are included in your User Test, we
suggest that you return the instrument to its default configuration
immediately after starting to program User Tests—Main Menu, Softkey
4 (Default Config.), Softkey 1.
68
Operation Manual
POWER-ON OPTIONS
Chap. 7
GN Otometrics
USER TEST PROGRAMMING
7.1
ZODIAC 901 Middle-Ear Analyzer