To Exit Setup
Test parameters that may be altered vary according to which test mode
is currently selected: e.g., there is no provision for changing pump speed
in reflex measurements, nor for adjusting the
y
-axis in tympanome-
try—only relevant parameters may be changed.
Setup may be exited simply by pressing the
Setup
pushbutton again.
Setup, More (Version
4.x Only)
In Tymp Auto Sweep and ETF-I modes, selecting
More
changes the sub-
menu choices to: optional use of a measurement pressure level a selected
interval above or below Middle-Ear Pressure (MEP), and Leak Detect on
or off (Version 4.
x
). Press
Setup
again to return to the function currently
selected.
In Reflex testing,
Setup
,
More
enables the addition of Leak Detect on or
off (Version 4.
x
).
Here follows an overview of the parameters that may be altered in tym-
panometry, reflex and threshold test modes. To change any test parame-
ter in Setup, press the corresponding
Softkey
repeatedly until the
required choice is shown in the panel above it. Alternatively, hold the
corresponding
Softkey
down and turn the
Rotary Knob
(in either direc-
tion).
5.2.1
TYMPANOMETRY (INCL. ETF-I)
Pump Speed
Pump Speed:
The settings available are: 50, 100, 200, 400 daPa/sec
and AFAP (as fast as possible).
This parameter is not available in Manual Tymp.
Scaling
Y-axis Scaling
(Compliance):
The following settings are available: 0.3, 0.6, 0.9, 1.5,
3.0, 4.5 ml or Auto. In Auto, the minimum possible
scaling for the measured curve is automatically
selected.
Averaging
Averaging:
On/Off (Manual Tymp. only).
Min. Pressure, Max.
Pressure
Min. Pres.,
Max. Pres.:
Minimum Pressure: –100, –150, –200, –250, –300, –350,
–400, –450, –500, –550, –600 daPa.
Maximum Pressure: 100, 150, 200, 250, 300, 350, 400
daPa.
Operation Manual
57
GN Otometrics
SUPPLEMENTARY FUNCTIONS
Chap. 5
ZODIAC 901 Middle-Ear Analyzer
SETUP
5.2
Fig.5.2
Tymp. Sweep Setup
Содержание Madsen ZODIAC 901
Страница 1: ...901 Middle Ear Analyzer User Manual Doc no 7 25 1800 11 Part no 7 25 180 0459 ...
Страница 6: ......