4.1.1
MANUAL
Manual
This test permits the user to operate the instrument manually. Both
Compliance and Volume tympanograms may be generated. Pressure
may be altered and volume adjusted to zero at any point in the test.
Softkey 1
Press
Main Menu
and use
Softkey 5
to select Advanced Mode (unless
this requirement has been changed using the ZODICON
TM
software).
Press
Tymp.
and then
Softkey 1
to access the screen shown in Fig.4.2.
“Vol.” shows the cavity volume measured at the pressure shown in the
center panel. “ECV” = Ear Canal Volume. “Comp.” = Compliance, i.e.,
the volume change relative to the ECV. Thus Vol. = ECV + Comp. ECV
is shown only when Comp. is selected. The vertical scales are in millili-
ters; 1 ml = 1 cc.
Software Version 3.x
and Higher Only
NOTE: The Tymp. screens of software versions lower than 3.
x
do not
include fields displaying calculated values for: MEP, SC, Grad., TW
nor Dir.
After a satisfactory Tymp. measurement has been made, and Comp. is
selected, a screen similar to that shown in Fig.4.3, containing the meas-
ured tympanogram, will be seen.
Operation Manual
31
GN Otometrics
MANUAL MODE
Chap. 4
ZODIAC 901 Middle-Ear Analyzer
TYMPANOMETRY
4.1
Fig.4.2
Tymp. sub-menu displayed when
Tymp. is selected in Advanced
ModeTymp. Manual before
measurement
Fig.4.3
Tymp. Manual after
measurement