4.3.3
AIR CONDUCTION
Air Conduction
Thresholds
This test mode enables the operator to use TDH 39 contra headphones
for determining air conduction thresholds. It also allows a bone-
conductor to be used for performing Gellé’s test, to investigate the mo-
bility of the ossicles, and especially of the stapes.
Softkey 3
Press
Main Menu
and use
Softkey 5
to select Advanced Mode. Press
Threshold
to display the Threshold Menu. Press
Softkey 3
from the
Threshold Menu to access the screen shown in Fig.4.22.
Tone Switch
Press the
Tone Switch
to present the stimulus.
Softkeys 1 / 3
The threshold values may be plotted on an audiogram on the dis-
play—just select frequency using
Softkey 1
and Sound Pressure Level in
dB using
Softkey 3
or the
Rotary Knob
.
Ext. Range
The range 0–100 dB may be extended by pressing the
Ext. Range
button
to 120 dB.
Softkey 2
Press
Softkey 2
to plot each threshold on the audiogram. Any values
may be entered; Fig.4.23 shows an example.
L/R Shift
When starting to test the patient’s other ear, remember to register the
change-over on the ZODIAC 901, otherwise the data for the first ear will
be overwritten. To change to the other ear, press
Alt. Select
and then
Softkey 4
.
Operation Manual
51
GN Otometrics
MANUAL MODE
Chap. 4
ZODIAC 901 Middle-Ear Analyzer
THRESHOLD TESTING
4.3
Fig.4.22
Air ConductionThreshold before
measurement
Fig.4.23
Air ConductionThreshold after
measurement