Curve For Each
Frequency
In Version 3.
x
and later, this test has been expanded so that the curve for
each frequency measured (Ipsi and Contra, for both ears) may be stored
and subsequently printed out.
Press
Main Menu
and use
Softkey 5
to select Advanced Mode (unless
this requirement has been changed using the ZODICON
TM
software).
Press
Reflex
to display the Reflex Menu.
Softkey 3
Press
Softkey 3
from the Reflex Menu to select Decay, and then press
Softkey 2
, or the
Tone Switch
, or the start button on the probe, to start
the test. The screen shown in Fig.4.17 will be displayed (the curve in this
illustration is typical of a normal ear).
Softkey 1
Press
Softkey 1
to set the desired stimulus type (500, 1000, 2000, 4000 Hz
or White Noise—note that White Noise is only available if Contra has al-
ready been selected.
Softkey 2
Softkey 2
starts or stops the test.
Softkey 3
Softkey 3
increases the intensity in increments of 5 dB (increments of 1,
2 and 10 dB are available via
Setup
); alternatively, use the
Rotary Knob
to increase or reduce stimulus intensity.
Ext. Range
The intensity range of 50 to 105 dB may be extended by pressing the
Ext.Range
button to 115 dB (or to 120 dB in Contra).
Softkey 4
Softkey 4
allows the selection of either Ipsi or Contra.
Softkey 5
Pressing
Softkey 5
displays the pump position graphically at the top of
the screen and enables you to adjust the pressure with the
Rotary Knob
as long as
Softkey 5
is kept pressed. When
Softkey 5
is released, the
pump-position scale disappears and the
Rotary Knob
reverts to control-
ling intensity. Note that, if the MEP function has been selected in Meas-
urement Setup, the pressure will be automatically changed by the
selected level before the reflex measurement is started.
L/R Shift
When starting to test the patient’s other ear, remember to register the
change-over on the ZODIAC 901, otherwise the data for the first ear will
44
Operation Manual
MANUAL MODE
Chap. 4
GN Otometrics
REFLEX TESTING
4.2
ZODIAC 901 Middle-Ear Analyzer
Fig.4.17
Reflex Decay