7.2 SYSTEM SETUP
Setup
Changes made to parameters using System Setup will be preserved in
the non-volatile memory of the ZODIAC 901 when the power is
switched off, and will remain in effect unless altered again subsequently
using the System Setup facility. To enter this mode, switch off the instru-
ment, and power it on again with
Setup
held down. Keep it held down
until the screen shown in Fig.7.6, System Setup, appears.
Softkey 1
Pressing
Softkey 1
scrolls through the languages available for the text
on the Liquid Crystal Display and printouts. Alternatively, hold
Softkey
1
down and turn the
Rotary Knob
clockwise or anticlockwise. Earlier
versions of the ZODIAC 901 allow a choice of English, Italian, Spanish
and German. Version 2.03 and higher also include French.
Softkey 2
Hold down
Softkey 2
and turn the
Rotary Knob
to adjust the sound
level of the beeper.
Softkey 3
Hold down
Softkey 3
and turn the
Rotary Knob
to adjust the contrast.
Softkey 4
Press
Softkey 4
to scroll through the different baud rates (300, 600, 1200,
2400, 4800, 9600)—the baud rate determines the speed of data transfer
wnen communicating with an external computer. Alternatively, hold
down
Softkey 4
and turn the
Rotary Knob
clockwise or anticlockwise.
Selecting Baud Rate
for External
Communications
IMPORTANT!
The baud rate must be set to the correct value (9600
baud) when communicating with MIDIMATE 622 or MateBaseI.
Operation Manual
69
GN Otometrics
POWER-ON OPTIONS
Chap. 7
ZODIAC 901 Middle-Ear Analyzer
SYSTEM SETUP
7.2
Fig.7.6
ZODIAC 901 System Setup