Disposal
·
There are no special requirements for disposal of the rubber eartips.
Probe Tip
·
The probe tip must be frequently checked for cerumen (ear wax)—even
a partial blockage will affect the accuracy of your measure- ments. Ce-
rumen may be removed by means of the supplied metal pin (cleaning
wire)—see below.
Do NOT attempt to clean any of the tubes in the probe tip without
first removing the attached plastic tubes as described below.
Procedure For
Cleaning Probe Tip
1.
First detach the probe tip from the 3 plastic tubes.
2.
Use the supplied metal pin (cleaning wire) to push out cerumen from
each metal tube in the probe tip.
3.
Re-attach the 3 plastic tubes to the probe tip.
2.5 FUSE REPLACEMENT
Voltage Select/ Fuse
Replacement
TO REPLACE A FUSE OR TO CHANGE VOLTAGE:
WARNING!
Before replacing a fuse or changing voltage, first switch off
instrument and disconnect from mains power supply!
1.
Switch off the instrument and disconnect from mains.
2.
Remove the two Phillips screws at the back of the top panel.
3.
Tilt the top panel forwards and remove.
Voltage Select
4.
The voltage selector is located next to the power switch (see Fig. 2.3 on
next page). 115 V or 230 V may be selected.
Fuse Replacement
5.
Two identical fuses are located beside the voltage selector (see Fig. 2.3). The
current of the fuses varies according to the voltage, see below:
Component
Voltage
Fuses
Type
Dimensions
F301/F302
115 V
400 mA
Slow Blow
5 x 20 mm
F301/F302
230 V
200 mA
Slow Blow
5 x 20 mm
Operation Manual
11
GN Otometrics
INSTALLING THE ZODIAC 901
Chap. 2
ZODIAC 901 Middle-Ear Analyzer
FUSE REPLACEMENT
2.5