A. APPENDIX: TECHNICAL SPECIFICATIONS
Standards
Audiometric Headset:
ISO 389, ANSI S3.6
Impedance:
IEC 1027, ANSI S3.39
Patient Safety:
EN 60601-1, class I, type B
Compliance Measuring System
Probe Tone:
226 Hz at 85 dB SPL
Total Range:
0.1 ml to 8.0 ml
Reflex Ranges:
0 to 75
m
l
0 to 150
m
l
0 to 225
m
l
0 to 300
m
l
0 to 375
m
l
Auto scaling and definition of positive,
negative or both types of axis
Display Range
Tympanogram:
0 to 1.5 ml
0 to 3.0 ml
0 to 4.5 ml
Air Pressure System
Range:
200 to -400 daPa
Ex400 to -600 daPa
Accuracy:
±
10% or
±
10 daPa (whichever is greater)
Air Pump:
Plunger type; stepper motor, digitally controlled.
Speeds:
as fast as possible (AFAP)
400 daPa/sec
200 daPa/sec
100 daPa/sec
50 daPa/sec
Manual speed control
Air Release:
Mechanical safety release valve set to
+600 and -800 daPa
Manual & automatic air release
Acoustic Reflex
Contralateral
Stimulation:
Pure tones: 0.5, 1, 2, 4 kHz
Frequency Accuracy
±
0.5%
White Noise
Step size: 1, 2, 5, 10 dB
Normal range: 50-100 dB HL
Extended range: 50-120 dB HL
Operation Manual
83
GN Otometrics
A
APPENDIX: TECHNICAL SPECIFICATIONS
ZODIAC 901 Middle-Ear Analyzer