5.2.3.2
FAST SCREENING
Stimulus Increments
Stimulus
Increments:
The following settings are available: 1, 2, 5, 10 dB.
Threshold Step
Threshold Step:
The following settings are available: 1, 2, 5, 10, 20, –1,
–2, –5, –10, –20 dB
MaxMeas
MaxMeas:
Selects the maximum number of measurements per
frequency from 1 to 16.
Y–Axis
Y
–Axis:
it is possible to select between Pos(itive), Neg(ative),
Pos/Neg and Auto.
If Auto is selected, the instrument will automatically
select whichever
Y
–axis displays the measured curve
in the greatest detail.
Use MEP
Use MEP
(Press.C.):
Selects whether automatic adjustment of MEP
(middle-ear pressure) is on or off. In Version 3.
x
,
Pressure Control includes: Manual, Auto, MEP+5,
+10, +15, +20, +25, –25, —20, –15, –10, –5 daPa. When
Auto is selected, an automatic offset to MEP is
calculated on the basis of Gradient and TW.
Leak Detect On/Off
Leak Detect:
On or Off
5.2.3.3
AIR CONDUCTION
Stimulus Increments
Stimulus
Increments:
The following settings are available: 1, 2, 5, 10 dB.
Pump On or Off
Pump:
Switches pump on or off (see Gellé’s Test, 4.3.3).
5.3 PRINTER
External Printer
The ZODIAC 901 Middle-Ear Analyzer is equipped with a thermal
printer as standard—it may, however, be supplied without the built-in
printer. Whether you have a built-in printer or not, your ZODIAC 901
has a parallel port on its rear panel for connection to an external printer.
Default installed in ZODIAC 901 is the HP generic Laserjet printer
driver which can be used together with any printer where PCL may be
selected. A Centronics cable for connection to a printer may be supplied
as an optional accessory. Other drivers for external printer may be de-
fined by means of ZODICON™, e.g. Epson FX, HP Deskjet 500 and IBM
Proprinter.
MIDIMATE 622
In addition, ZODIAC 901 may be connected via the RS232C Data Inter-
face serial port to the MIDIMATE 622 Audiometer, to allow the MIDI-
MATE 622 to print out audiograms on ZODIAC 901’s built-in printer or,
60
Operation Manual
SUPPLEMENTARY FUNCTIONS
Chap. 5
GN Otometrics
PRINTER
5.3
ZODIAC 901 Middle-Ear Analyzer
Содержание Madsen ZODIAC 901
Страница 1: ...901 Middle Ear Analyzer User Manual Doc no 7 25 1800 11 Part no 7 25 180 0459 ...
Страница 6: ......