Keratron™ Installation and operating manual
OPTIKON 2000
Cod. 161501EN
8-1
2014-04-23
Rev.0
8
TROUBLESHOOTING GUIDE
8.1 TROUBLESHOOTING
This section describes the messages that identify operating situations of the
Keratron™ that have not been described previously in the manual.
Symptom or Message
Significance
Action
Keratron
™
does not turn on
The unit is not
correctly powered:
the power cable is
not connected or the
mains fuse has
burned out.
Check the power cable.
Replace the fuse (see
section § 10.3.1).
The space between
the mires cone’s
photoelectric cells
during the first
calibration step
(Figure 7-4) is not
free.
Repeat the calibration
with the target (Figure
7-2) positioned as
shown in Figure 7-4. If
the error message
persists, contact the
Optikon 2000 S.p.A.
Service Department.