Keratron™ Installation and operating manual
OPTIKON 2000
Cod. 161501EN
7-2
2014-04-23
Rev.0
7.1 TOPOGRAPHY CALIBRATION [CALIBRATION]
Selecting [Calibration] on the
touchscreen
[Calibration] accesses the section
regarding the calibration of the instrument for topography.
The provided aluminum disk (
calibration target
shown in Figure 7-2) and
calibration sphere
, Figure 7-3, are necessary for this operation.
Figure 7-2: Calibration target
Figure 7-3: Calibration sphere
The entry screen requires that the calibration target be positioned as shown in Figure
7-4. Pressing the [OK] button causes the machine to perform an initial reading of the
photoelectric cells that are free of obstacles.
If the operator presses the [Abort] key, the
Keratron™ returns to the initial screen,
exiting calibration.