Keratron™ Installation and operating manual
OPTIKON 2000
Cod. 161501EN
7-6
2014-04-23
Rev.0
Figure 7-10: Final calibration screen
NOTE:
In the calibration steps during the various acquisitions of the calibration sphere, it is
advisable to first use the
joystick
until the middle OK arrow (3 in Figure 7-7) lights up
in a stable manner, and THEN press the button to acquire the image
.
Avoid reaching
the OK position with the button already pushed down.
IMPORTANT:
It is good practice to repeat the calibration daily when you turn on the Keratron
™
.
However, it is indispensable to perform the calibration after the instrument has been
installed or any time it is moved or cleaned.
7.2 SETTINGS
Selecting on
touchscreen
function
Settings
allows you to access a window that lists
which firmware and software versions are present on
Keratron™ or configure several
options of the unit itself.
In the
About
folder, in addition to the unit's serial number
,
you will find listed the
version and
checksum
of the software components of the user interface (Onda.exe
and Onda.dll). The same information is listed for the resident
firmware
in the unit’s
electronics. Optikon 2000 S.p.A. can be consulted as the software designer of all of
these applications.