Keratron™
Installation and operating manual
OPTIKON 2000
Cod. 161501EN
5-8
2014-04-23
Rev.0
5.3.2 GUIDANCE AND MANUFACTURER'S DECLARATION
–
ELECTROMAGNETIC IMMUNITY
The
Keratron™ unit intended for use in the electromagnetic environment specified
below. The customer or user of the
Keratron™ should assure that it is used in such
an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic Environment
– Guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
· ±6 kV contact
· ±8 kV air
· ±6 kV contact
· ±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical/Fast
transient/burst IEC
61000-4-4
· ±2 kV for power supply ines
· ±1 kV for input/output ines
· ±2 kV for power supply lines
· ±1 kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital environment
Surge
IEC 61000-4-5
· ±1 kV differential mode
· ± 2kV common mode
· ±1 kV differential mode
· ± 2kV common mode
Mains power quality should be that of a
typical commercial or hospital environment
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
· 0% UT for 0.5 cycles
· 40% UT (60% dip in UT) for 5
cycles
· 70% UT (30% dip in UT) for
25 cycles
· 0% UT for 5 sec
· <5% UT (>95% dip in UT) for
0.5 cycles
· 40% UT (60% dip in UT) for 5
cycles
· 70% UT (30% dip in UT) for
25 cycles
· <5% UT (>95% dip in UT) for
5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the
Keratron™ requires
continued operation during power mains
interruptions, it is recommended that the
Keratron™ uninterruptible power suplí or a
battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
Not applicable
The
equipment
does
not
contain
any
magnetically
sensitive components.
Power frequency magnetic field should be at
level characteristic of a typical location in a
typical commercial or hospital environment.
Conducted RF IEC
61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3V/m
da 80 MHz a 2.5 GHz
3 Vrms
3V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the
Keratron™, including cables,
than the recommended separation distance
calculated from he equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d= 1.2 ÖP
d= 1.2 ÖP 80MHz to 800 MHz
d= 2.3 ÖP 800MHz to 2.5 GHz
where
P
is the maximum output power
rating to the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the
recommended separation distance in
metres (m)
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the compliance
level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: