C200
H
ARDWARE
Instruction Manual C200 Part1 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 7
Accuracy:
±0.25% FSD, ±5% of reading (whichever is the greater)
Resolution:
0.1% at 10mA, 0.05% at 20mA
Maximum load resistance:
750
Ω
at 20mA
Configuration:
Can be assigned to either measured variable or either sample
temperature
Access to Functions
Direct keypad access:
Measurement, maintenance, configuration, diagnostics and
service
functions
Performed without external equipment or internal jumpers
Control Function – C210 and C220 only
Controller Type:
P, PI, PID (configurable)
Output:
Can be assigned a maximum of two relays, two analog outputs
or one of each
Analog:
Current output control (0 to 100%)
Time proportioning cycle time:
1.0 to 300.0s, programmable in increments of 0.1s
Pulse frequency:
1 to 120 pulses per minute, programmable in increments of
1 pulse per minute
Controller action:
Reverse, direct or bi-directional (programmable)
Proportional band:
0.1 to 999.9%, programmable in increments of 0.1%
Integral action time (Integral reset):
1 to 7200s, programmable in increments of 1s (0 = Off)
Derivative:
0.1 to 999.9s programmable in increments of 0.1s, available
only for single set point control
Auto/Manual: User-programmable
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...