C200
S
OFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 97
Resistance Span (25
Ω
)
Set the cell simulator to 25
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Resistance Span (2.5
Ω
)
Set the cell simulator to 2.5
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Resistance Span (0.5
Ω
)
Set the cell simulator to 0.5
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Zero (1K
Ω
)
Set the temperature simulator to 1k
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Span (1K5)
Set the temperature simulator to 1.5k
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Continued on next page.
A: Res 2.5 ohm
2.500
2.500
A: Res 25 ohm
25.00
25.00
A: Res 0.5 ohm
0.500
0.500
A:T.Zero (1K0)
1000
1000
A: Ref. Checking
A:T.Span (1K5)
1500
1500
Ohms
Calib
Ohms
Calib
Ohms
Calib
Ohms
Calib
Ohms
Calib
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...