C200
S
OFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 72
5.2.4.3 Configure
CIP Status Alarm
Notes.
•
This section is applicable only if A1: Type is set to Dosing – see Section 5.2.4.1.
•
Alarm 5 (dual input analyzers only with option board fitted and analog features enabled – see Section 6.2.3)
can be configured as a CIP Status alarm for Sensor B, therefore this section applies also to Alarm 5.
•
When Alarm 1 (and/or Alarm 4) is configured as a CIP Interface alarm, the error messages shown in Table
7.2.1 are displayed in response to the events described.
•
If Alarm 2 is configured as a CIP Status alarm, its associated relay is energized/de-energized and its LED lit
and extinguished in response to the events described in Table 7.2.1.
Configure Alarm 2
See Section 5.4.1.
Alarm 2 Type
Note
. The CIP Status alarm type is displayed only if A1: Type is set to Dosing –
see Section 5.2.4.1.
Use the
UP
and
DOWN
keys to select the CIP Status alarm type.
When A2: Type is set to CIP Status:
• and the measured conductivity has not exceeded 90% of the Alarm 1 set point
value by the end of the initial charge time period, Alarm 2 relay energizes,
Alarm 2 LED is lit and a Long Charge error message is displayed (see Table
7.2.1).
• and the measured conductivity has not exceeded the Alarm 1 set point value
by the end of the dosing time period, Alarm 2 relay energizes, Alarm 2 LED is
lit and a Long Dose error message is displayed (see Table 7.2.1).
• and the measured conductivity exceeds the Alarm 1 set point value plus the
deviation set point value, Alarm 2 relay energizes, Alarm 2 LED is lit and a
High Alarm error message is displayed (see Table 7.2.1).
• and the measured conductivity falls below the Alarm 1 set point value minus
the deviation set point value, Alarm 2 relay energizes, Alarm 2 LED is lit and a
Low Alarm error message is displayed (see Table 7.2.1).
Off, Status or CIP Status selected – return to top of page.
Alarm selected – see Section 5.2.4.1.
Config. Alarm 2
-----
A2: Type
-----
A2: Assign
Config. Alarm 3
Config. Alarm 2
CIP
Status
Alarm
Off
Status
A1: Type
set to
Dosing
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...