C200
S
OFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 89
Factory Settings Access Code
Enter the required code number, between 0000 and 19999, to gain access to the
factory settings. If an incorrect value is entered, access to subsequent frames is
prevented and the display reverts to the top of the page.
Calibrate Sensor A
Note.
The values in the display lines for sensor calibration are shown as
examples only – the actual values obtained will differ.
Sensor B calibration (dual input analyzers only) is identical to
Sensor A calibration.
See page 91.
Return to Operating Page – see Section 2.1.2.
Calibrate Input for Sensor A ?
If calibration is required select Yes otherwise select No.
Note.
To abort calibration, press the
DOWNSCROLL
key again at any time
before calibration is complete – see next page.
Millivolt Zero
Set the millivolt source to –1000mV.
The display advances automatically to the next step once a stable and valid
value is recorded.
Note.
The upper 6-segment display shows the measured input voltage. Once the
signal is within range the lower 6-segment display shows the same value and
Calib is displayed to indicate that calibration is in progress.
Millivolt Span
Set the millivolt source to +1000mV
The display advances automatically to the next step once a stable and valid
value is recorded.
Temperature Zero (100R)
Set the temperature simulator to 100
Ω
The display advances automatically to the next step once a stable and valid
value is recorded.
Continued on next page.
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A:mV Zero (-1V)
-1000
Calib
A:mV Span (+1V)
mV
Calib
A:T.Zero (100R)
Ohms
Calib
Cal. Sensor B
Monitoring pH
-1000
mV
A: T.Zero (1K0)
1000
1000
100.0
100.0
Cal. Output 1
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...