2.1.2 Operating
page
2.1.2.1 Single Input pH
Measured Values
pH.
Temperature.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see
Set Min Slope
in the
CONFIG. SENSORS
page – Section 5) and 105% is displayed. If the value is outside
these limits, check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Alarm 3 set to
Wash
(
A3: Type
, Section 5) – see Section 2.1.2.6
Alarm 3 not set to
Wash
.
Monitoring pH
7.00
pH
24.4
Deg.C
Wash Function
Monitoring pH
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...