C200
S
OFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 70
5.2.4.2 Configure
CIP Interface Alarm
Notes.
•
This section is applicable only if A1: Type is set to Dosing – see Section 5.2.4.1.
•
Alarm 4 (dual input analyzers only with option board fitted and analog features enabled – see Section 6.2.3)
can be configured as a CIP Interface alarm for Sensor B, therefore this Section applies also to Alarm 4.
•
When Alarm 1 (and/or Alarm 4) is assigned as a CIP Interface alarm, the error messages shown in Table
5.2.4.3on page 68 are displayed in response to the events described.
Alarm 1 Failsafe
Select Yes to enable failsafe action, otherwise select No.
Alarm 1 Action
Select the alarm action required, High or Low.
Alarm 1 Set Point
Set the alarm set point to the required value within the sensor's measurement
range
Initial Charge
Select Yes to enable an initial charge sequence. The initial charge sequence
commences when A: Initial Charge is set manually to Yes in the Operating Page
(Section 2.2.2.1) and continues for the initial charge time period set below.
Initial Charge Time
Set the initial charge period, between 1 and 30 minutes in 1 minute increments.
Continued on next page.
A1: Action
Low
High
-----
A1: Failsafe
No
Yes
-----
Initial Charge
Init. Charge Time
30
Mins
A1: Setpoint
1000
mS/cm
A1: Type
set to
Dosing
No
Yes
Dosing
Yes
No
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...