C200
S
OFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH
•
D-45356 Essen
•
Germany
page 39
Calibrate Buffer 2 (Two-Point Calibration only)
Immerse Sensor A in the second buffer solution.
pH and Adjust are shown alternately on the upper display line. Use the and
keys to set the displayed pH value to the temperature-corrected pH value of the
chosen solution (see the datasheet provided with the solution).
Press the
DOWNSCROLL
key to initiate calibration.
Note.
To abort calibration, press the
DOWNSCROLL
key again at any time
before calibration is complete – see previous page.
Measured pH value from the last successful calibration.
Measured sensor output in milivolts.
When the measured sensor output stabilizes, press the key to accept the
calibration.
Cal. Accepted is shown for 2 seconds to confirm that the calibration has been
accepted, the display then advances automatically to the next frame.
The display then changes for 2 seconds to show the temperature-corrected
buffer value in the upper display line, then advances automatically to the next
frame.
Calibration Message
See Table 4.1 for details of calibration messages.
Slope Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope –
Section 5.1.3) and 105% is displayed. If the value is outside these limits, check
the electrode system.
pH check value.
Displayed as an additional indication of electrode system condition; 7pH is the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Sensor B (dual input analyzers only) calibration is identical to
Sensor A calibration.
Probe Type for Sensor B (dual input analyzers only) set to Redox or
ORP and Enable Cals. set to No (Section 5.1.3) – return to top of
page.
Alter Sec. Code not set to zero (Section 5.1.8) – see Section 5.1.1
Alter Sec. Code set to zero (Section 5.1.8) – see Section 5.1.2.
A: Immerse Buf.2
9.18
pH
20.0
Deg.C
A: Calibration
-----
Passed
A: Slope & Check
100.0
%
7.00
pH
A: Cal. Buffer 2
9.22
pH
A: Buffer Method
set to
Man 1-Pt
A: Buffer Method
set to
Man 2-Pt
9.22
105
mV
pH
Press button
when stable
A: Cal. Accepted
9.22
pH
SECURITY CODE
CONFIG. DISPLAY
Sensor Cal. A
Sensor Cal. B
SENSOR CAL.
Содержание CONTROL 200
Страница 95: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 73 ...
Страница 128: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 106 ...
Страница 130: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 108 8 3 PID Controller ...
Страница 131: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 109 ...
Страница 133: ...C200 SOFTWARE Instruction Manual C200 Part2 A1 doc optek Danulat GmbH D 45356 Essen Germany page 111 ...