High-Penetration Software Option
The 45MG high-penetration software option coupled with low-frequency single
element transducers (down to 0.5 MHz) allows you to make thickness, material sound
velocity, and time-of-flight measurements on materials such as composites, fiberglass,
plastic, rubber, and cast metals that are difficult or impossible to measure using
standard ultrasonic instruments. The M2008 [U8415001] transducer is a special low-
frequency transducer used to measure the thickness of thick fiber-reinforced polymer
(FRP) and composite materials.
With a M2008 transducer only, press
[2nd F]
,
[CAL ZERO] (Do ZERO)
at any time to
auto-adjust the zero offset and compensate for temperature changes in the delay line.
To use the high-penetration software option with the M2008 transducer
1.
Ensure that the high-penetration software option is activated (see “Activating
Software Options” on page 68 for details).
2.
Connect the M2008 transducer to the T/R 1 connector at the top of the 45MG.
3.
Press
[2nd F]
,
[FREEZE] (XDCR RECALL)
.
4.
In the menu, select
DEFAULT HP SINGLE ELEMENT
.
5.
In the
DEFAULT HP SINGLE ELEMENT
screen, highlight the default transducer
setup for the M2008 (
DEFP1-0.5-M2008
), or any custom setup that uses the M2008
transducer.
6.
Press
[MEAS]
to return to the measurement screen with the recalled transducer
setup.
7.
Wipe the couplant off the transducer tip.
8.
Press
[2nd F]
,
[CAL ZERO] (Do ZERO)
.
9.
Perform the material sound velocity and zero calibration (see “Calibrating the
Instrument” on page 55).
6.7
Datalogger Option
This section describes how to use the 45MG internal datalogger to organize your data.
NOTE
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...