DMTA-10022-01EN, Rev. C, January 2015
Technical Specifications
211
Calibration
Single or two-point calibration using a test block
Velocity can be entered manually.
Display response time
Normal (4 Hz) or fast (up to 20 Hz) for dual element
transducers
Adjustable: 4 Hz, 8 Hz, 16 Hz, and Max (up to 20 Hz) for
single element transducers
Number of pixels to display a
waveform
320 × 240 pixels
Table 22 Environmental rating specifications
Parameter
Value
IP rating
Designed for IP67
Explosive atmosphere
MIL-STD-810G, Section 511.5, Procedure I
Shock test
MIL-STD-810G, Section 516.6, Procedure I
Vibration test
MIL-STD-810G, Section 514.6, Procedure I
Drop test
MIL-STD-810G, Section 516.6, Procedure IV- Transit Drop
Table 23 Measurement specifications
Parameter
Value
Measurement modes
Standard dual element
: Time between excitation pulse and
first back-wall echo using dual element transducer.
Dual Echo-to-Echo
: Time between successive back-wall
echoes using dual element transducers.
THRU-COAT
: Time between excitation pulse and first back-
wall echo while ignoring or displaying a coating thickness.
Mode 1
: Time between excitation pulse and first echo
following a blank period using contact transducers.
Mode 2
: Time between the interface echo and the first back-
wall echo. Normally used with delay line or immersion
transducers.
Mode 3
: Time between a pair of back-wall echoes following
the interface echo. Normally used with delay line or
immersion transducers.
V-path correction
Automatic, dependent on transducer type
Table 21 Other EN15317 specifications
(continued)
Parameter
Value
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...