DMTA-10022-01EN, Rev. C, January 2015
Chapter 9
158
•
Save a live thickness reading while the B-scan is running.
•
Save any reviewed thickness reading on a frozen B-scan.
•
Save all thickness readings for one B-scan screen, along with the minimum or
maximum thickness readings for a held B-scan.
•
Save an entire B-scan history of up to 10000 thickness readings, the minimum or
maximum thickness readings, and the held B-scan.
To save a live thickness reading while the B-scan is running
Press
[SAVE]
.
To save any reviewed thickness reading on a frozen B-scan
1.
While the B-scan is running, press
[FREEZE]
to enter the review mode.
2.
Use the
[
]
and
[
]
keys to review any thickness on the frozen B-scan.
3.
Press
[SAVE]
to save the thickness value for the review marker position to the
datalogger.
To save the minimum or maximum thickness reading for a frozen B-scan
1.
Set
B-SCAN FREEZE MODE
to
SHOW MINIMUM
or
SHOW MAXIMUM
.
The gage displays the minimum or maximum thickness reading with the
corresponding waveform.
2.
While the B-scan is running, press
[FREEZE]
to enter the review mode.
3.
Press
[SAVE]
while the minimum or maximum thickness reading is displayed.
To save the entire B-scan (B-scan history)
1.
In the
B-SCAN
screen (see Figure 9-4 on page 154), set
SAVE KEY DATA
to
THK + B-SCAN
.
2.
While the B-scan is running, or when it is frozen, press
[SAVE]
.
The
Save B-Scan history?
message appears on the help text bar.
3.
Select
Yes
to save the entire B-scan history, including the Minimum or Maximum
thickness with the corresponding waveforms.
OR
Select
No
to save the current B-scan screen and the waveform.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...