DMTA-10022-01EN, Rev. C, January 2015
Chapter 10
184
Figure 10-17 Saving custom setups
5.
In the
ACTIVE
screen, review the setup parameters.
6.
Press
[MEAS]
to return to the measurement screen.
10.10 Quickly Recalling a Custom Setup for Single Element Transducers
Normally, you can change the setup for a custom application by pressing
[RECALL XDCR]
to select the appropriate setup in the list of available setups), and
then pressing
[MEAS]
. This process is adequate for typical setups that are changed
infrequently. However, when it is necessary to change between two or more custom
setups on a more frequent basis, use the quick setup recall function instead.
When the quick setup recall function is activated, a simple keypad shortcut can be
used to quickly switch to any of the first four custom setups for single element
transducers.
To activate the quick setup recall function
1.
In the measurement screen, press
[SETUP]
, and then select
MEAS
.
2.
In the
MEAS
screen, set
QUICK SETUP RECALL
to
ON
.
3.
Press
[MEAS]
to return to the measurement screen.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...