DMTA-10022-01EN, Rev. C, January 2015
Chapter 10
166
Additionally, when establishing a custom setup for a specific application, it is
essential to verify the performance on reference standards representing the materials
and thickness range to be measured.
To create a custom setup for a single element transducer
1.
Connect the single element transducer to the 45MG (see “Setting Up the
Transducer” on page 51).
2.
Press
[2nd F]
,
[FREEZE] (XDCR RECALL)
.
3.
In the menu, select the
CUSTOM SINGLE ELEMENT
.
4.
In the
CUSTOM SINGLE ELEMENT
screen, select the desired custom setup
location (
SE-USER-n
) in which you want to save the custom setup.
To reduce the number of changes needing to be made to parameter values, you can
also select an existing single element setup with parameter values that are close to the
desired setup.
5.
In the
ACTIVE
screen (see Figure 10-1 on page 166):
a)
Set
DET MODE
to the desired detection mode (see “Detection Modes” on
page 169 for details).
Figure 10-1 The ACTIVE screen for the setup of a single element transducer
TIP
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...