DMTA-10022-01EN, Rev. C, January 2015
Managing Communications and Data Transfer
189
If you experience problems establishing communication between the 45MG and a
remote device, consider using the 45MG communication reset function to return all
communication parameters to their default values (see “Resetting the Communication
Parameters” on page 196), and then reconfigure only those communication
parameters that are needed.
11.3 Exchanging Data with a Remote Device
It is possible to exchange 45MG data with a remote device such as a computer.
Data transmitted from the 45MG remains in the 45MG internal memory.
11.3.1
Exporting a File to the Memory Card (Datalogger Option Only)
The 45MG has the ability to export files from the internal memory to the external
microSD memory card. Files can be exported in CSV (comma-separated variables),
text (space delimited), or survey (used by the GageView interface program) format. A
microSD card reader can then be used to open these files in Microsoft Excel, or any
other program, directly on your computer. Survey files can be imported into
GageView from the microSD card reader.
To export files to the external memory card
1.
Ensure that a microSD memory card is inserted in the designated slot under the
battery door of the 45MG (see Figure 1-4 on page 21).
2.
If you inserted the microSD memory card when the 45MG was already turned on,
turn off and restart the 45MG to force it to recognize the presence of the memory
card.
3.
In the measurement screen, press
[FILE]
, and then select
EXPORT
.
TIP
NOTE
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...