DMTA-10022-01EN, Rev. C, January 2015
Appendix A
212
Measurement resolution
Selectable from the keypad:
LOW
: 0.1 mm (0.01 in.)
STD
: 0.01 mm (0.001 in.)
HI
: 0.001 mm (0.0001 in.) with the High Resolution option.
Not all resolutions are available for all measurement modes.
Material sound velocity range 0.508 mm/µs to 18.699 mm/µs (0.0200 in./µs to 0.7362 in./µs)
Material sound velocity
resolution
0.001 mm/µs (0.0001 in./µs)
Alarm set point range
0.00 mm to 500.00 mm (0.00 in. to 20.00 in.)
Table 24 Datalogger specifications
Parameter
Value
Storage capacity
475000 thickness readings or
20000 waveforms with thickness readings
ID number length
1 to 20 characters
File name length
1 to 32 characters
File formats
Incremental
Sequential (defined by starting and ending ID number)
Sequential with custom points
2-D grid
Boiler
External memory card
microSD memory card
2 GB maximum capacity
Table 25 Typical ranges and default setups for single element
transducers
a
Setup name
Transducer
Typical measurement range
DEFM1-20.0-M116
M116
Steel: 0.250 mm to 8.000 mm (0.020 in. to 1.500 in.)
DEFM1-10.0-M112
M112
Steel: 0.760 mm to 250.000 mm (0.030 in. to 10.000 in.)
DEFM1-10.0-M1016
M1016
Steel: 0.760 mm to 250.00 mm (0.030 in. to 10.000 in.)
DEFM1-5.0-M110
M110
Steel: 1.00 mm to 380.00 mm (0.040 in. to 15.000 in.)
DEFM1-5.0-M109
M109
Steel: 1.00 mm to 500.00 mm (0.050 in. to 20.000 in.)
DEFM1-2.25-M106
M106
Steel: 2.00 mm to 635.00 mm (0.080 in. to 25.000 in.)
Table 23 Measurement specifications
(continued)
Parameter
Value
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...