DMTA-10022-01EN, Rev. C, January 2015
Maintaining and Troubleshooting the 45MG
205
7.
When
DUAL XDCR TEST
is selected, in the
DUAL XDCR TEST
screen (see
a)
Press
[MEAS]
to start the test that measures the time of flight through each
side of a dual element transducer.
The measured parameter values appear.
b)
Interpret the
Tx
and the
Rx
values as follows:
•
Similar values indicate a normal dual element transducer.
•
A difference between the values indicates that the delay line wear is not
the same for each element.
•
A missing value indicates that a cable is broken, or that the element is not
functioning.
c)
Take note of the calculated
ZERO VALUE
.
d)
Press
[ENTER]
to terminate the dual element transducer test.
Figure 12-3 The DUAL XDCR TEST screen
12.6 Performing the Software Diagnostic Test
The software diagnostic (
SW DIAG
) function generates an error log documenting all
errors that occurred during instrument operation. Olympus uses this information to
troubleshoot the 45MG operating software.
To access the software diagnostic
1.
In the measurement screen, press
[SETUP]
, and then select
SP MENU
.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...