DMTA-10022-01EN, Rev. C, January 2015
Chapter 3
36
Figure 3-5 Other elements — Waveform option enabled
The loss of signal (
LOS
) appears and the thickness value is cleared when the 45MG no
longer detects ultrasonic echoes (see Figure 3-6 on page 36).
Figure 3-6 The loss-of-signal (LOS) indicator
In the rare case that the returning signal from the transducer is beyond the upper
receiver-voltage limit, the
LOS
indicator will be replaced by the
SAT
flag for
indication of saturation. This situation can normally be corrected by lowering the
pulser voltage.
Zoom indicator
Rectification indicator
Echo detection marker
Delay value
Mode indicator (min/max)
and value
Freeze indicator
Alarm indicator
Gain
Range value
[WAVE ADJ]
parameter
Loss-of-signal (
LOS
)
indicator (or
SAT
in the
case of saturation)
No thickness value found
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...