DMTA-10022-01EN, Rev. C, January 2015
Chapter 8
144
8.2
Configuring System Parameters
The
SYSTEM
screen, allows you to configure many 45MG system parameters.
To configure system parameters
1.
In the measurement screen, press
[SETUP]
, and then select
SYSTEM
.
The
SYSTEM
screen appears (see Figure 8-2 on page 144).
Figure 8-2 The SYSTEM screen
2.
Set
BEEPER
to
ON
or
OFF
(see “Setting the User Interface Language and Other
System Options” on page 43 for details).
3.
Set
INACTIVE TIME
to
ON
or
OFF
(see “Setting the User Interface Language
and Other System Options” on page 43 for details).
4.
In the
LANGUAGE
text box, select the desired user interface language (see
“Setting the User Interface Language and Other System Options” on page 43 for
details).
5.
Set
RADIX TYPE
to the desired character (period or comma) for separation of the
integer and decimal numbers.
6.
Set
SAVE KEY DATA
to save only the thickness measurement (
THICKNESS
), or
both the thickness and the waveform (
THI WF
).
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...