DMTA-10022-01EN, Rev. C, January 2015
Basic Operation
51
5. Basic Operation
The following sections describe the basic operation for the 45MG ultrasonic thickness
gage.
5.1
Setting Up the Transducer
The 45MG operates with a full line of single element (optional software) and dual
element transducers. The 45MG automatically recognizes standard D79X dual
element transducers, and automatically loads the appropriate predefined setup. The
predefined setup contains ultrasonic velocity for the stainless steel step block
supplied with the instrument. With dual element transducers, you need to perform a
transducer zero compensation.
For Single Element or High Penetration software options and single element
transducers, you need to manually recall an appropriate setup. The 45MG is factory
shipped with default conditions for the purchased transducers set up using an
approximate sound velocity for the stainless steel test block provided with the
instrument. The default conditions are selected to facilitate instrument usage for your
applications.
To set up the transducer
1.
Plug the transducer into the transducer connector(s) on the top of the 45MG case
(see Figure 5-1 on page 52), using only T/R 1 for a single element transducer.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...