DMTA-10022-01EN, Rev. C, January 2015
Chapter 5
58
5.2.2
Test Blocks
The 45MG comes with a cylindrical stainless steel test block with two thicknesses.
Two precisely known test-block thicknesses can be used to perform the material
sound velocity and the zero calibrations.
Precision step test blocks are also often used when more than two known thicknesses
are needed (see Figure 5-8 on page 58).
Figure 5-8 Example of a five-step test block
When performing the material sound velocity and zero calibrations, you must use a
test block with the following characteristics:
•
Made from the same material as the part to be tested.
•
Has two or more precisely known thicknesses.
•
Has one part as thin as the thinnest section of the parts to be tested in order to
perform the zero calibration. The surface condition should be similar to that of the
tested parts. Rough surfaces generally reduce the accuracy of measurements, but
simulating actual surface conditions on the calibration block can help to improve
results.
•
Has one part as thick as the thickest section of the parts to be tested, in order to
perform the material sound velocity calibration. The front and back surfaces
should be smooth and parallel.
•
Be at the same temperature as the samples to be measured.
5.2.3
Transducer Zero Compensation
Perform a transducer zero compensation by pressing
[2nd F]
,
[CAL ZERO]
(
Do ZERO
) whenever the zero indicator (
Do--
) appears. This procedure should also
be performed when the dual element transducer temperature has changed.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...