DMTA-10022-01EN, Rev. C, January 2015
Software Options
79
•
Adjust the waveform range and delay.
— With the Single Element transducer option (see “Custom Setups for Single
Element Transducers” on page 165):
•
Adjust the pulse-receive parameters (TVG gain and blanks)
•
Create custom single element transducer setups.
•
Adjust the waveform range and delay.
•
Auto Zoom: This mode automatically adjusts the range and delay so that the
measured echo is always on-screen.
•
The waveform rectification: Enables the user to switch the waveform between Full
wave, Half+, Half
−
, and RF display.
•
Waveform Fill: Enables the user to show the filled-in waveform. This feature is
not available for RF rectification mode.
•
Assigns the
[SAVE]
key to:
Save only the thickness value.
OR
Save the thickness and waveform (only available when the Datalogger software
option is also purchased).
Figure 6-7 Standard display
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...