DMTA-10022-01EN, Rev. C, January 2015
Chapter 4
44
Figure 4-1 The SYSTEM screen
3.
Press
[MEAS]
to return to the measurement screen.
4.
Turn off the 45MG, and then turn it back on to activate the language change.
4.2
Selecting the Measurement Units
The 45MG can be set to show thickness measurements in inches or millimeters.
To set the measurement units
1.
In the measurement screen, press
[SETUP]
, and then select
MEAS
.
2.
In the
MEAS
screen, set
UNIT TYPE
to
INCH
or
MILLIMETER
.
3.
Press
[MEAS]
to return to the measurement screen.
4.3
Setting the Clock
The 45MG has a built-in date and time clock. You can set the date and the time and
select their format. The 45MG saves all measurement values with their acquisition
date.
To set the clock
1.
In the measurement screen, press
[SETUP]
, and then select
SP MENU
.
2.
In the
SP MENU
(see Figure 4-2 on page 45), select
CLOCK
.
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...