PRECESSION ELEPHANT 2 Scan Head 2
Abort criterion
Galvo
o
ffset dx
and/or
dy
larger than ± 1 %
–
OR
–
Limit reached at least at one of the micrometer screws
Condition
The laser beam must not be clipped at the nozzle and has to be centered to the
beam exit.
not OK
OK
19. If the above condition is
not
OK then adjust either via the InScript software
–
OR
–
mechanically via the adjustable nozzle at the focusing optics unit.
TIP
Adjust one axis first and then the other.
InScript software:
a. In the InScript software in the
Navigator
view, click
Pens
default
head
.
b. In the
Node Properties
view on the
A Data
tab set
Galvo offset dx
and/or
dy
until the condition above is met.
Adjustable nozzle:
•
Turn the micrometer screws until the condition above is met.
20. Abort the job.
21. If the abort criterion below is met then repeat section
on page
else
proceed with section
on page
42