PRECESSION ELEPHANT 2 Scan Head 2
Table 9: VCSEL parameters
Class 3B laser product according to DIN EN 60825-1:2008-05
P
0
= 100 mW
Maximum average power accessible during
normal
operation with scan
head
housing closed
P
0
= 400 mW
Maximum average power accessible during
service
operation with scan
head
housing opened and vision unit removed
λ
= 850 nm
Wavelength
Tool
•
uEye Cockpit
software [
] for viewing the images delivered by optional
through-the-lens vision camera
Procedure
1. If your scan head does not have a through-the-lens vision camera then proceed
with section
on page
2. In document [
], look-up how the
Vcsel
device is configured and used.
3. View the image delivered by the through-the-lens vision camera with the
uEye
Cockpit
software [
].
4. In the InScript software in the
Navigator
view in the controller’s
Devices
subtree,
select the
vcsel
device.
5.
CAUTION
Invisible laser radiation
could cause damage to eye or skin.
a. Avoid eye or skin exposure to direct or scattered radiation.
b. In the
Node Properties
view, set the following variables as needed.
i. Set
Channel 1
to
Enable
.
ii. Slowly turn up
Duty cycle
and watch the image delivered by the
camera. Turn Duty cycle up until the scan field is well lit and the
image is not saturated yet.
6. Proceed with subsection
on page
60