PRECESSION ELEPHANT 2 Scan Head 2
No. Label / description
Action
①
Cooling water in- or outlet;
Water quality:
DI-water, filtered, particle size < 100 µm;
Pressure drop: 3 to 4 bar (300 to 400 kPa);
Volume flow: 1.2 to 2.2 l/min;
Inlet temperature: max. 30 °C (303 K);
Recommend tubing material:
polyetherurethane; D 6 mm, d 4 mm;
The cooling unit is DI-water-proofed with
corrosion resistant types of steel.
Connect with tube to cooling
water supply; see also
section
on page
②
SLAVE LINK / XY4-100
This connector is irrelevant for the PRECESSION
ELEPHANT scan head, as the scan head already
uses all data channels and thus a second scan
head cannot be controlled. Furthermore the
scan head cannot be controlled by the XY4-100
protocol.
None
③
DATA I/O connector
Connect with CAT-5 patch
cable to the SCAN HEAD
DATA A connector of the ASC
system controller, see
④
Cooling water out- or inlet;
see also no. 1
Connect with tube to cooling
water supply; see also
section
on page
21