Section 9: Specifications
9- 2
Operating Altitude
-390m to +3,048m (-1,280 ft. to
+10,000 ft.)
Operating Relative Humidity
15 to 95%, noncondensing
9.5
ALARMS
Adult Alarm Limit Range
Systolic Blood Pressure
60–250 mmHg
Diastolic Blood Pressure
40–220 mmHg
Mean Blood Pressure
45–235 mmHg
Pulse Rate
30–180 bpm
Neonatal Alarm Limit Range
Systolic Blood Pressure
40–130 mmHg
Diastolic Blood Pressure
20–90 mmHg
Mean Blood Pressure
35–105 mmHg
Pulse Rate
30–240 bpm
9.6
FACTORY DEFAULTS
Factory Default Alarm Limit Settings
Adult
Neonatal
Systolic Blood Pressure Upper Limit:
240 mmHg
120 mmHg
Systolic Blood Pressure Lower Limit:
70 mmHg
50 mmHg
Diastolic Blood Pressure Upper Limit:
210 mmHg
80 mmHg
Diastolic Blood Pressure Lower Limit:
50 mmHg
30 mmHg
Mean Blood Pressure Upper Limit:
225 mmHg
95 mmHg
Mean Blood Pressure Lower Limit:
55 mmHg
45 mmHg
Pulse Rate Upper Limit:
170 bpm
230 bpm
Pulse Rate Lower Limit:
40 bpm
40 bpm
Other Factory Default Settings
Operating Mode
Adult-Pediatric
Audible Alarm Volume
Level 5
Audible Alarm Silence Period
60 seconds
Alarm Silence Reminder
On
Measurement Interval
Off (Manual)
Measurement Completion Tone
On
Pressure Calibration Algorithm
2 (Adjusted)
Initial Inflation Pressure (Adult-Pediatric)
180 mmHg
Initial Inflation Pressure (Neonatal)
120 mmHg