Section 6: Disassembly Guide
6- 5
6.3.3 Replacing the Hose Fitting
To replace the hose fitting:
1.
Perform step 1 in paragraph 6.3.2.
2.
Disconnect the pump hose from the fitting on the side panel.
3.
Loosen and remove the retaining nut.
4.
Pull fitting out of the side-panel molded housing.
5.
Install the new fitting, applying LocTite (removable thread locker 242) or
equivalent on the hose fitting and retaining nut, as illustrated in Figure
6-4.
6.
Ensure that the hose is located as indicated in Figure 6-4 to avoid
crimping when the monitor is reassembled. Connect the hose to the
fitting.
6.3.4 Removing the UIF Board
When replacing the UIF PCB, ensure that the Instrument Identification on
the bottom of the unit (see Figure 6-2) is replaced with the one provided with
the replacement PCB. To remove the UIF board:
1.
Disassemble the monitor as indicated in paragraph 6.3 and 6.3.2, steps 1
through 3.
2.
Remove the five screws securing the board to the chassis, as shown in
Figure 6-5.
3.
Lift up, carefully prying up on the left side where J3 is connected to the
top docking connector, which is mounted to the chassis.
4.
Disconnect the Display board from the UIF board.