Section 9: Specifications
9- 3
9.7
PERFORMANCE
Range
Adult-Pediatric Measurement Range:
Systolic Blood Pressure
60–250 mmHg
Diastolic Blood Pressure
40–220 mmHg
Mean Arterial Blood Pressure
45–235 mmHg
Pulse Rate
40–200 bpm
Neonatal Measurement Range:
Systolic Blood Pressure
40–130 mmHg
Diastolic Blood Pressure
20–90 mmHg
Mean Arterial Blood Pressure
35–105 mmHg
Pulse Rate
40–240 bpm
Measurement Accuracy
Adult-Pediatric:
Pulse Rate Accuracy: Displayed pulse rates in the range of 40–200 bpm are
accurate to within
±
2 bpm or 2%, whichever is greater.
Pressure scale range in the adult-pediatric mode is 10–300 mmHg. Values in
the pressure scale range will be displayed while the N-3100 is inflating and
deflating the cuff. When the cuff pressure is within this range, the pressure
indicated by the N-3100 is within
±
3 mmHg or
±
2%, whichever is greater, of
the actual cuff pressure.
Note:
Accuracy is not specified for displayed values outside of the stated
ranges.
Neonatal:
Pulse Rate Accuracy: Displayed pulse rate in the range of 40–240 bpm are
accurate to within
±
2 bpm or 2%, whichever is greater.
Pressure scale range in the neonatal mode is 5–150 mmHg. Values in the
pressure scale range will be displayed while the N-3100 is inflating and
deflating the cuff. When the cuff pressure is within this range, the pressure
indicated by the N-3100 is within
±
3 mmHg or
±
2%, whichever is greater, of
the actual cuff pressure.
Note:
Accuracy is not specified for displayed values outside of the stated
ranges.