Section 4: Configuration and Service Modes
4- 10
4.3.1 Menu Item 1: Software Version Report
This report identifies the software versions of the UIF PCB and blood
pressure module.
1.
From the service mode steady state, select menu item 1 by pressing the
UP or DOWN ARROW buttons until “1” appears in the SYSTOLIC
display. Press the UPPER ALARM LIMIT button. The UIF software
version number is displayed in the left-hand character of the DIASTOLIC
display. The UIF module major revision number is displayed in the next
two characters of the DIASTOLIC display. The UIF module minor
revision number is displayed in the MEAN/PULSE RATE display.
2.
Press the UP ARROW button to view the blood pressure module version
in the SYSTOLIC display.
3.
Press the LOWER ALARM LIMIT button to return to the service mode
steady state.
4.3.2 Menu Item 2: Lamp Test Mode
This test verifies that indicators and front-panel lamps are functional.
1.
From the service mode steady state, select menu item 2 by pressing the
UP or DOWN ARROW buttons until “2” appears in the SYSTOLIC
display. Press the UPPER ALARM LIMIT button.
2.
Initially, all indicators will light. Press and hold the UP ARROW button
to light each display segment, decimal, and indicator, in a sequential
scanning pattern to verify that each lamp works. After each lamp has lit,
all lamps will turn off.
Note:
The POWER-ON indicator is not tested with this procedure. It
can be verified by turning the monitor on and off.
3.
Press and hold the DOWN ARROW button to reverse the firefly pattern.
4.
Press the LOWER ALARM LIMIT button to return to the service mode
steady state.
4.3.3 Menu Item 3: Button Test Mode
This test verifies proper operation of individual buttons and button
combinations.
1.
From the service mode, select menu item 3 by pressing the UP or DOWN
ARROW buttons until “3” appears in the SYSTOLIC display. Press the
UPPER ALARM LIMIT button.