6- 1
SECTION 6: DISASSEMBLY GUIDE
6.1
Introduction
6.2
Fuse Replacement
6.3
Monitor Disassembly
6.4
Lithium Battery Replacement
6.5
Reassembly
6.1
INTRODUCTION
The N -3100 can be disassembled down to all major component parts,
including:
•
PCBs
•
lithium batteries
•
cables
•
function keys
•
chassis enclosures
Tools required for disassembly:
•
small, Phillips-head screwdriver
•
medium, Phillips-head screwdriver
•
needle-nose pliers
WARNING : Before attempting to open or disassemble the N-3100,
disconnect the AC power cord from the N-3100.
Caution: Observe ESD (electrostatic discharge) precautions when
working within the unit.
If the bottom cover is removed while disassembling the monitor, do
not connect the SPS Power Supply to the monitor while the power
supply is plugged into AC power. Instead, first connect the power
supply to the monitor, then connect the power supply to AC power.
Failure to do so may result in damage to the monitor.
Note:
Some spare parts have a business reply card attached. When you
receive these spare parts, please fill out and return the card.
6.2
FUSE REPLACEMENT
See Figure 6-1. To replace the fuse on the lower docking connector board:
1.
Disconnect the N-3100 from the SPS power supply. If you are operating
in the stacked configuration, remove the N-3000 from the stack.
2.
Set the N -3100 upside down, facing you.
3.
Loosen the two screws on the left end of the bottom cover.
4.