Section 4: Configuration and Service Modes
4- 3
4.2.1 Adult-Pediatric/Neonatal Mode Default
The operating mode that the monitor is in when you exit the configuration
mode (adult-pediatric or neonatal), becomes the power-on default operating
mode.
4.2.2 Systolic Upper Alarm Limit
1.
From configuration mode steady state, press the UPPER ALARM LIMIT
button once. The current default value is displayed in the SYSTOLIC
display.
2.
To change the upper alarm limit value, press and hold the UP or DOWN
ARROW button on top of the monitor. You cannot decrease the value
lower than the set lower alarm limit.
3.
To place the monitor back into configuration mode steady state, allow the
display to time out (3 seconds).
4.2.3 Systolic Lower Alarm Limit
1.
From configuration mode steady state, press the LOWER ALARM LIMIT
button once. The current default value is displayed in the SYSTOLIC
display.
2.
To change the lower alarm limit value, press and hold the UP or DOWN
ARROW button on top of the monitor. You cannot increase the value
higher than the set upper alarm limit.
3.
To place the monitor back into configuration mode steady state, allow the
display to time out (3 seconds).
4.2.4 Diastolic Upper Alarm Limit
1.
From configuration mode steady state, press the UPPER ALARM LIMIT
button twice. The current default value is displayed in the DIASTOLIC
display.
2.
To change the upper alarm limit value, press and hold the UP or DOWN
ARROW button on top of the monitor. You cannot decrease the value
lower than the set lower alarm limit.
3.
To place the monitor back into configuration mode steady state, allow the
display to time out (3 seconds).
4.2.5 Diastolic Lower Alarm Limit
1.
From configuration mode steady state, press the LOWER ALARM LIMIT
button twice. The current default value is displayed in the DIASTOLIC
display.
2.
To change the lower alarm limit value, press and hold the UP or DOWN
ARROW button on top of the monitor. You cannot increase the value
higher than the set upper alarm limit.
3.
To place the monitor back into configuration mode steady state, allow the
display to time out (3 seconds).