Section 4: Configuration and Service Modes
4- 2
Note:
While changing limits, there must be some user interaction with the
monitor within a 3-second period or the monitor returns to
configuration mode steady state operation.
Table 4-1 lists the default settings that can be configured and the respective
buttons to press to access the settings. Methods used to change the default
settings are detailed in paragraphs 4.2.1 through 4.2.13.
Table 4-1: Configuration Mode Menu
Default Setting
Buttons to Press While in
Configuration Mode Steady State
Systolic Upper Alarm Limit
Press UPPER ALARM LIMIT button once
Systolic Lower Alarm Limit
Press LOWER ALARM LIMIT button once
Diastolic Upper Alarm Limit Press UPPER ALARM LIMIT button twice
Diastolic Lower Alarm Limit
Press LOWER ALARM LIMIT button twice
Mean Upper Alarm Limit
Press UPPER ALARM LIMIT button three
times
Mean Lower Alarm Limit
Press LOWER ALARM LIMIT button three
times
Pulse Rate Upper Alarm
Limit
Press UPPER ALARM LIMIT button four
times
Pulse Rate Lower Alarm
Limit
Press LOWER ALARM LIMIT button four
times
Alarm Volume
Press and hold ALARM SILENCE button
Alarm Silence Duration
Press ALARM SILENCE button and within 3
seconds press UP or DOWN ARROW button
Auto Measurement Cycle
Interval
Press CYCLE button
Configuration Menu
Press UPPER and LOWER ALARM LIMIT
buttons simultaneously
Adult-Pediatric/Neonatal
Power-up Mode
Press NEW PATIENT/NEONATAL button
twice within 2 seconds
Note:
To change neonate default alarm limits, first enter the neonatal
mode from configuration mode steady state by pressing the NEW
PATIENT/NEONATAL button twice within 2 seconds. Then change
the desired limit using the same method as for adult default limits.