Section 3: Performance Verification
3- 2
3.
Observe the monitor front panel. With the N-3100 in STANDBY (turned
off), press the ON/STANDBY button. In order to successfully complete
the self-test, the monitor must perform the following sequence.
a.
The monitor emits three consecutively higher-pitched beeps.
b.
All indicators light momentarily as illustrated in Figure 3-1. Verify
that “8.8.8.” is shown in the SYSTOLIC, DIASTOLIC, and
MEAN/PULSE RATE displays.
Note:
Slight differences in the brightness of display segments and
indicators are normal and acceptable.
SYS
mmHg
DIA
mmHg
MAP
mmHg
AUTO
NEO
MAN
Figure 3-1: Self-Test Display
c.
All displays (except the POWER ON indicator) turn off briefly.
d.
Each segment in the digital displays lights individually in sequence
in a scanning, or firefly, test pattern and all indicators light
momentarily.
e.
The sequence of steps b and c will be repeated until all internal tests
are complete.
Note:
The number of times the sequence of steps b and c is
repeated is dependent on the amount of information stored
in trend memory. Typically, you will see two cycles of the
sequence.
f.
A single, 1-second beep is produced and all displays again illuminate
momentarily, indicating the automatic Power-On Self-Test is
complete.
g.
The POWER ON indicator and the MANUAL MODE indicator are
illuminated. A “–” is displayed in each window of the SYSTOLIC,
DIASTOLIC and MEAN/PULSE RATE displays. This is referred to
as the normal mode steady state.
SYS
mmHg
DIA
mmHg
MAP
mmHg