Section 3: Performance Verification
3- 5
11. From the normal mode steady state, press and release the UPPER
ALARM LIMIT button three times. Verify that the monitor emits a
single beep each time you press the button. Verify that the
MEAN/PULSE RATE display indicates an alarm limit value of “225” for
a few seconds and verify that the other displays show dashes at the top
while the “225” is displayed.
12. From the normal mode steady state, press and release the UPPER
ALARM LIMIT button three times. Press and hold the DOWN ARROW
button until “55.” is displayed in the MEAN/PULSE RATE display.
13. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button three times. Verify that the monitor emits a
single beep each time you press the button. Within 3 seconds, verify that
the MEAN/PULSE RATE display indicates the default alarm limit value
of “55” and verify that the other displays show dashes at the bottom
while the “55” is displayed.
14. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button three times. Within 3 seconds, press and hold the
DOWN ARROW button until “55.” is shown in the MEAN/PULSE RATE
display. Verify that “55.” is the lowest value to which the lower alarm
limit can be adjusted.
15. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button three times. Within 3 seconds, press and hold the
UP ARROW button and verify that “55.” is the highest value to which
the lower alarm limit can be adjusted.
16. From the normal mode steady state, press and release the UPPER
ALARM LIMIT button four times. Verify that the monitor emits a single
beep each time you press the button. Verify that the PULSE RATE
indicator is lit and the MEAN/PULSE RATE display indicates the
default alarm limit value of “170” for a few seconds and verify that the
other displays show a dash at the top of each display character while the
“170” is displayed.
17. From the normal mode steady state, press and release the UPPER
ALARM LIMIT button four times. Press and hold the DOWN ARROW
button until “40.” is shown in the MEAN/PULSE RATE display.
18. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button four times. Verify that the monitor emits a single
beep each time you press the button. Within 3 seconds, verify that the
PULSE RATE indicator is lit and the MEAN/PULSE RATE display
indicates the default alarm limit value of “40” and verify that the other
displays show dashes at the bottom while the “40” is displayed.
19. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button four times. Within 3 seconds, press and hold the
DOWN ARROW button until “30.” is displayed in the MEAN/PULSE
RATE display. Verify that “30.” is the lowest value to which the lower
alarm limit can be adjusted.
20. From the normal mode steady state, press and release the LOWER
ALARM LIMIT button four times. Within 3 seconds, press and hold the
UP ARROW button and verify that “40.” is the highest value to which
the lower alarm limit can be adjusted.